COMPACT test chambers (16, 60 ,110 and 200 l) DY - DY T

Small & Powerful

ACS Compact Test Chambers provide high efficiency in limited space. Small footprint, easy installation, low consumption, reduced noise level and reach basic configuration, including the observation window, make these chambers the ideal choice for any Research and Development Laboratory looking for compact units to test small components and products in the electronic and automotive fields.

They are equipped with the powerful KeyKratos Plus touchscreen control system, based on PID control through PLC, having the following advanced features:

  • Memory capacity to 1000 cycles with 350 segments (on Secure Digital).
  • Unlimited real time recording of temperature and humidity (only for 110 and 200 l) versus time (log on Secure Digital).
  • Recordings in CSV format (Comma Separated Value) for easy export to Excel® from Secure Digital to Pendrive via USB interface placed on the front panel (program files are easily convertible into graphic format).
  • Messages on scheduled maintenance activities.
  • A trend chart for recorded variables available with a scale from -100/+200.
  • Synoptic graphs for the chamber, low stage refrigeration and high stage refrigeration.

ACS Benchtop Chambers can provide a temperature range from -40° to +130°C (16 l) or -70° to +180°C (60, 110 and 200 l) and humidity range from 10% RH to 95% RH (110 and 200 l). All the models only require a single-phase electrical supply, except for the 200 l model.

Potential-free contact for switching-off test specimens, UV lamp and additional shelves are available as options for 110 and 200 l models.

16, 60 and 110 l chambers can easily sit on a laboratory benchtop and the 110 l model can be supplied with an optional supporting table with wheels .

Multi-chamber remote control is available as option together with WinKratos S/W.

Models and technical specifications:

Model DY16 T DY16 T(C) DY110 (C) ¹ DY200 (C) ¹
Useful capacity (l) 16 59.5 110 206
Internal dimensions (mm) Width 310 350 548 601
Depth 230 340 447 541
Height 206 500 447 634
External dimensions (mm) Width 465 630 877 927
Depth 541 970 1080 1379
Height 685 1180 1400 1794
Temperature range (°C) Basic -35 to +130 -40 to +180 -40 to +180 -40 to +180
C model   -70 to +180 -70 to +180 -70 to +180
Temperature fluctuation (K) ±1 ±0.1 to ±0.3 ±0.1 to ±0.3 ±0.1 to ±0.3
Temperature changing rate
Heating 4+5 (K/min)
Basic (-40/+180°C) 4.5 3 3.2 4
C model (-70/+180°C)   3 3.2 4
Temperature changing rate
Cooling 4+5 (K/min)
Basic (+180/-40°C) 3.5 3 4 4.5
C model (+180/-70°C)   3 2.8 3
Humidity range (%) (τ=-3/+93°C) 2     10 to 95 10 to 95
Temperature range for climatic test (°C)     10 to 95 10 to 95
Humidity fluctuation (%)     ±1 to ±3 ±1 to ±3
Maximum thermal load (W) 5 Basic T=+25°C 250 250 350 2300
C model T=+25°C   250 500 1500
Rated power (kW) Basic 0.7 2.3 3 6
C model   3.9 3.7 7
Rated current absorption (A) Basic 4 11 16 10
C model   15 16 12
Weight (kg) Basic 60 210 350 6 485
C model   230 360 6 545
Sound pressure level dB(A) 3 Basic 52 57 52 53
C model   59 52 59
Supply voltage (Vac) 230V
±10%/50Hz/1
+ G
400V
±10%/50Hz/3
+ N+ G
1. for Temperature only version add the suffix T - 2. τ=+4/+93°C for continuous test - 3. measured a 1 m distance in front of the unit, free field measurement - 4. according to IEC 60068-3-5 and IEC 60068-3-6 - 5. The performance data refer to +22°C ambient temperature, 230V or 400V nominal volatage, without specimen - 6. Value without supporting table